Product Details
CIBIO® Single Panel Drug Test Cassette
Rapid Single Drug Test Cassette is a rapid, screening test for the qualitative detection of single drug and drug metabolites in human urine at specified cut off levels.
The tests are intended for In Vitro Diagnostics use. They are intended for prescription use including point of care sites and over-the-counter use.
Product Name | Single Panel Drug Test Cassette | |
Brand Name | CIBIO® or OEM | |
Format | Cassette | |
Methodology | Colloidal Gold | |
Kit Contents | 1. Test Cassette | |
2. Dropper | ||
3. Desiccant | ||
4. Instruction for use | ||
Assay Type | Qualitative Detection | |
Specimen | Urine | |
Specification | 4.0 mm | |
Certificates | ISO 13485, CE | |
Accuracy | Over 99% | |
Storage Temperature | 4-30 degree centigrade | |
Shelf Life | 2 years ( 24 months ) | |
Delivery time | 15-30 days | |
OEM Services | Pouch, Box, Label, Insert | |
Loading Port | Guangzhou or Shenzhen |
Product Pictures:
Drug Tests Available And Cut-off Levels |
|||
1 | Amphetamine (AMP) | 500/ 1000 ng/mL | |
2 | Barbiturates (BAR) | 300 ng/mL | |
3 | Benzodiazepines (BZO) | 300 ng/mL | |
4 | Buprenorphine (BUP) | 10 ng/mL | |
5 | Cocaine (COC) | 150/ 300 ng/mL | |
6 | Marijuana (THC) | 25/ 50 ng/mL | |
7 | Methamphetamine (MET) | 500/ 1000 ng/mL | |
8 | Methadone (MTD) | 300ng/mL | |
9 | Methylenedioxymethamphetamine- Ecstasy(MDMA) |
500ng/mL | |
10 | Morphine 300 (MOP) | 300ng/mL | |
11 | Opiate 2000 (OPI) | 2000ng/mL | |
12 | Oxycodone (OXY) | 100g/mL | |
13 | Phencyclidine (PCP) | 25g/mL | |
14 | Propoxyphene (PPX) | 300ng/mL | |
15 | Tri-cyclic Antidepressants (TCA) | 1000ng/mL | |
16 | Ethyl -glucuronide(ETG) | 300 or 500ng/mL | |
17 | Fentanyl(FYL) | 200ng/mL | |
18 | Synthetic Marijuana (K2) | 50ng/mL | |
19 | 6-Monoacetylmorphine(MAM) | 10ng/mL | |
20 | Nicotine(COT) | 200ng/mL | |
21 | Tramadol(TRA) | 200ng/mL | |
22 | Ketamine (KET) | 1000ng/mL |
How To Do The Test?
Test must be at room temperature 59-86°F (15-30°C).
1. Remove a Testing Device from the foil pouch by tearing at the notch and place it on a level surface.
2. Holding a sample dropper vertically, add 3 drops of the urine specimen to the sample well (with an arrow marked).
3. Read the result in five minutes. Positive results may sometimes appear as early as one minute but you should wait the full 5 minutes before determining a negative result.
Reading The Results:
Preliminary positive (+)
A rose-pink band is visible in the control region, but no colored band appears in the test region. This indicates a preliminary positive result for this specific drug. Preliminary positive results must be sent to a lab for a more reliable test.
Negative (-)
A rose-pink band is visible in the control region and the test region. This indicates that the drug is not present in the sample or is at a level that cannot be detected by this test.
Invalid
If a color band is not visible in the control region, or a color band is only visible in the test region, the test is invalid. You should repeat the test with a new test dipcard. If the test still shows an invalid result, please contact us at 1-864-538-8007. When calling, be sure to provide the lot number for the test.
Note: There is no meaning to the colored line intensity or width. Any visible line is considered to be a line.
Certain lines may appear lighter or thinner than other lines.
ANY COLORED LINE VISIBLE IN THE TEST “T” REGION, NO MATTER HOW DARK OR FAINT, SHOULD BE INTERPRETED AS A NEGATIVE RESULT AS LONG AS THE CONTROL LINE IS ALSO PRESENT.
IMPORTANT:
This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug test result, particularly when preliminary positive results are indicated.